How to Calibrate OMNIR 4-20mA analog outputs
This describes how to match the OMNIR’s 4-20mA transmitter to the customer’s 4-20mA receiver. See the wiring schematic for additional details.
Step |
| |
---|---|---|
1 | Open the OMNIR windows software. Click Diagnostics, then IO Check. Confirm pausing the system. |
|
2 | Click the Calibrate button next to Channel 0 or 1. |
|
3 | Click SET LOW VALUE and read the mA value received on your system. Enter the value read in the box. |
|
4 | Click SET HIGH VALUE and read the mA value received on your system. Enter the value read in the box. |
|
5 | Click ACCEPT CALIBRATION |
|
6 | Enter 8.0 in the box shown in the figure. Ensure your system reads the same value. If the values are not close enough, repeat steps 1-5 . |
|
7 | If necessary, repeat the above steps for any additional channels. |
|