Distance Calibration Procedure
Purpose
Use this guide to calibrate a distance sensor on a transverse pad.
Step-by-step guide
STEP |
| |
---|---|---|
1 | Ensure the distance sensor is clean prior to starting. |
|
2 | Open the Distance sensors configuration under the Setup menu. | |
3 | Press the Calibrate… button next to the sensor to calibrate. | |
4 | Place the calibration block on the sensor so the 0.000” side is blocking the distance sensor. The distance sensor hole should be completely covered. | |
5 | Press the Continue button in the software. |
|
6 | Rotate the calibration block so the 1.250” section is straddling the distance sensor hole. | |
7 | Press the Finish button in the software. |
|
8 | New factor and offset values should appear. |
|
9 | Press Accept in the software. |
|
10 | Place a board over the distance sensor, navigate to the Static display, and confirm the distance value is accurate. |
|