Distance Calibration Procedure

Purpose

Use this guide to calibrate a distance sensor on a transverse pad.

Step-by-step guide

STEP

 

STEP

 

1

Ensure the distance sensor is clean prior to starting.

 

2

Open the Distance sensors configuration under the Setup menu.

3

Press the Calibrate… button next to the sensor to calibrate.

4

Place the calibration block on the sensor so the 0.000” side is blocking the distance sensor. The distance sensor hole should be completely covered.

5

Press the Continue button in the software.

 

6

Rotate the calibration block so the 1.250” section is straddling the distance sensor hole.

7

Press the Finish button in the software.

 

8

New factor and offset values should appear.

 

9

Press Accept in the software.

 

10

Place a board over the distance sensor, navigate to the Static display, and confirm the distance value is accurate.